Independent validation of four-dimensional flow MR velocities and vortex ring volume using particle imaging velocimetry and planar laser-Induced fluorescence. : Validation of 4D Flow using PIV and PLIF
PURPOSE:This study aimed to: (i) present and characterize a phantom setup for validation of four-dimensional (4D) flow using particle imaging velocimetry (PIV) and planar laser-induced fluorescence (PLIF); (ii) validate 4D flow velocity measurements using PIV; and (iii) validate 4D flow vortex ring volume (VV) using PLIF.METHODS:A pulsatile pump and a tank with a 25-mm nozzle were constructed. PIV